Laboratory surveillance of dengue in the municipality of São Paulo: challenges of agreement between diagnostic methods

Laboratory Surveillance of Dengue in the Municipality of São Paulo: Challenges of Agreement Between Diagnostic Methods

Luciano Marcondes de Oliveira; Gustavo Dantas Lobo

DOI: 10.22167/2675-6528-202603191

Article derived from a Course Conclusion Work (TCC), with content based on the student’s original work and adapted to the editorial format of the E&S Magazine with the support of the ResumeAI tool, an artificial intelligence solution developed by Instituto Pecege for textual synthesis and organization.

Summary

Dengue represents a major public health challenge in the municipality of São Paulo, demanding rapid and accurate laboratory diagnostics, encompassing methodologies for capturing distinct biological targets. The statistical agreement between ELISA-NS1 assays (commercial kit Bioclin, for NS1 antigen detection) and RT-qPCR (detection of RNA and identification of viral serotype), both used in the routine of the Zoonoses Laboratory (LABZOO) of the Municipality of São Paulo, was investigated, evaluating the influence of epidemiological, laboratory, and clinical variables. A retrospective analysis was performed with 1,146 tests from suspected dengue cases, in serum samples collected between January and October 2025. To overcome non-random sampling, agreement analysis (Kappa index) was applied, in addition to the non-parametric Mann-Whitney and Kruskal-Wallis tests for comparison between concordant and non-concordant results. The agreement between the laboratory methods varied according to the seasonal epidemiological behavior of dengue and the age group of the individuals. Concordant results showed higher absorbance values in the ELISA test and lower Ct values in RT-qPCR (higher viral load). The NS1 antigen demonstrated greater stability over the time of symptoms when compared to viral RNA. The low agreement of the commercial ELISA kit with the RT-qPCR assay constituted a limiting factor for the consistency of laboratory diagnosis. Therefore, the inclusion of rigorous analytical criteria in tender processes for the acquisition of ELISA kits, the strengthening of internal quality control, and the consolidation of an integrated diagnostic workflow between ELISA-NS1, RT-qPCR, and ELISA-IgM assays at LABZOO are recommended.

Keywords: Arbovirosis; Analytical Performance; ELISA-NS1; RT-qPCR; Association Tests.

1. Introduction

Dengue is an arbovirus disease caused by a virus of the genus Flavivirus, primarily transmitted by the bite of the Aedes aegypti mosquito. The pathology presents a broad clinical spectrum, ranging from undifferentiated febrile forms to severe cases with hemorrhagic manifestations and shock (Guzman et al., 2010). In Brazil, its epidemiological dynamics are marked by the simultaneous circulation of four viral serotypes (DENV-1 to DENV-4), a factor that increases the complexity of disease monitoring and control (Teixeira, 2008; Vicente et al., 2016).

In 2025, the State of São Paulo registered 871,939 cases of dengue, with 1,114 deaths (Center for Epidemiological Surveillance, 2025). In the same period, the municipality of São Paulo, which has an estimated population of 11.9 million inhabitants (Brazilian Institute of Geography and Statistics, 2025), accounted for 65,711 cases and 38 deaths (Coordination of Health Surveillance, 2025). This epidemiological scenario in a metropolis of high demographic density requires a robust laboratory structure, capable of processing significant volumes of samples to support both clinical management and vector control strategies. The Laboratory for the Diagnosis of Zoonoses and Vector-Borne Diseases (LABZOO), a reference for the municipal network’s diagnosis, processed 17,284 serum samples for dengue diagnosis in 2025. Of this total, 14,877 tests (86.1%) were for samples collected between the first and fifth day of symptoms for NS1 antigen research, while 2,407 (13.9%) were for IgM antibody research in samples collected from the sixth day onwards.

The LABZOO protocol for the acute phase is based on the detection of the NS1 glycoprotein via enzyme-linked immunosorbent assay (ELISA). In all reactive samples using the ELISA-NS1 method, the Reverse Transcription Polymerase Chain Reaction in Real Time (RT-qPCR) technique is applied for the detection of viral RNA and identification of the circulating serotype (Castro-Jorge, 2010; Secretaria Executiva de Atenção Básica, 2025). Although ELISA-NS1 is widely used for its agility and operational feasibility, its performance is influenced by variables such as viral load, interval between symptom onset and sample collection, epidemiological seasonality, and infecting viral serotype (Lorenz, 2025; Teoh et al., 2016).

With the introduction of the ELISA-NS1 Biolisa Bioclin commercial kit at LABZOO in 2025, a variation in the agreement between the ELISA method and RT-qPCR was observed, especially during periods of low dengue transmission. In this context, a high number of positive samples in ELISA showed a negative result in RT-qPCR. This disparity burdened the laboratory workflow due to the referral of samples for RT-qPCR, a methodology of higher cost and complexity, without corresponding molecular confirmation (Pillay et al., 2025). Altman and Bland (1994) described that the performance of diagnostic tests can vary depending on the epidemiological scenario, the nature of the biological material, and the viral load in the host.

The detection differences between RT-qPCR and ELISA-NS1 lie in the biochemical nature of their targets. While viral RNA exhibits intrinsic instability, being susceptible to rapid degradation by endogenous RNases and thermal variations during transport, the NS1 antigen possesses greater structural robustness, proving less sensitive to pre-analytical degradation processes (Duyen et al., 2011). Although RT-qPCR is widely used as the gold standard due to its high specificity, studies indicate distinct detection windows between viral RNA and the NS1 antigen (Erra et al., 2013; Tricou et al., 2011). This biological characteristic poses a relevant question to laboratory routine, as discordant results can represent either analytical specificity failures or reflections of different stages of viral kinetics (Peeling et al., 2010). The difficulty in distinguishing non-specific reactions from legitimate detections in phases of declining viral replication constitutes a relevant diagnostic challenge.

Therefore, the present investigation is justified by the need to optimize the laboratory flowchart for the diagnosis of dengue in the municipality of São Paulo. The objective of this study was to evaluate the statistical agreement between the ELISA-NS1 (Kit Bioclin) and RT-qPCR assays performed by LABZOO from January to October 2025, analyzing how epidemiological variables (seasonality and serotype), laboratory variables (absorbance intensity and viral load indicators), and patient characteristics (age group and symptom duration) influenced the agreement between the assays, aiming to provide technical subsidies for the optimization of the laboratory flowchart for the diagnosis of dengue in the municipality of São Paulo.

2. Material and Methods

The present study was characterized as a retrospective data analysis, observational in nature and quantitative in approach. The investigation was conducted at the Laboratory of Zoonoses and Vector-Borne Diseases (LABZOO), belonging to the Zoonoses Surveillance Division of the Municipality of São Paulo. All procedures were carried out in accordance with current ethical guidelines for the treatment of anonymized data.

The unit of analysis comprised a total of 1,146 tests, performed on serum samples collected from patients with clinical suspicion of dengue. The samples were obtained during the viremic phase, defined as the period from zero to five days after symptom onset, between January and October 2025. Participants were selected based on diagnostic test results.

The analyzed laboratory flow included samples that presented positive, negative, or inconclusive results in the enzyme immunoassay for the detection of the NS1 antigen (ELISA-NS1), using a commercial Bioclin kit. All these samples were subsequently subjected to the reverse transcription followed by real-time polymerase chain reaction (RT-qPCR) assay, according to the protocol of the Adolfo Lutz Institute (Barbosa, 1999), for etiological confirmation and identification of the viral serotype. RT-qPCR was established as the reference method for confirming the results.

Due to the laboratory workflow adopted, in which RT-qPCR was not systematically applied to all samples tested by ELISA in the network, the dataset analyzed did not represent a simple random sampling of the population. As signaled by Pepe (2003) and Zhou et al. (2011), the presence of conditional or convenience sampling directed the analyses towards the evaluation of agreement between the methods, as well as the verification of the influence of laboratory and epidemiological variables on this agreement.

The treatment, cleaning, and exploratory analysis of the data were executed in the Python language (version 3.12). For structured data manipulation, the Pandas library (McKinney, 2010) was used. The SciPy (Virtanen et al., 2020) and Statsmodels (Seabold and Perktold, 2010) libraries were employed for the application of statistical models.

The normality hypothesis of the distribution of continuous variables was evaluated using the Shapiro-Wilk test (Shapiro and Wilk, 1965). Given the absence of normality in the data distribution, the use of non-parametric statistical inference was chosen (Conover, 1999). For the comparison of continuous variables between two independent groups, the Mann-Whitney test was applied (Mann and Whitney, 1947).

When the analysis involved the comparison of three or more independent groups, the Kruskal-Wallis test (Kruskal and Wallis, 1952) was used. In situations where this test indicated a statistically significant difference, Dunn’s post hoc test (Dunn, 1964) was applied for pairwise comparisons, with Bonferroni correction (Dunn, 1961) to control the error rate. The association between categorical variables was evaluated by Pearson’s chi-squared test (Pearson, 1900).

In cases where the chi-squared assumptions were violated, characterized by reduced expected frequencies in the contingency table, Fisher’s exact test (Fisher, 1922) was used. The agreement between the categorical results of ELISA-NS1 and RT-qPCR was estimated by Cohen’s Kappa coefficient (Cohen, 1960), with interpretation based on the criteria of Landis and Koch (1977). The correlation between continuous or ordinal variables was assessed by Spearman’s coefficient (Spearman, 1904). In all analyses, a significance level of 5% (p = 0.05) was adopted.

3. Results and Discussion

The retrospective analysis of 1,146 tests from suspected dengue cases, collected between January and October 2025, revealed important demographic and temporal characteristics of the studied population. A slight predominance of the female sex was observed, representing 53.7% of the sample. The distribution of collections was markedly seasonal, with more than half of the samples concentrated in March and April, followed by May and June. This concentration reflects the peak of dengue transmission in the municipality of São Paulo, according to official data from the Coordination of Health Surveillance (COVISA, 2025), a common pattern in Brazilian urban epidemics, where the climatic conditions of the first and second quarters favor the proliferation of the vector (Teixeira, 2008).

The age range of participants demonstrated a wide variation, encompassing individuals from newborns to 96 years old, with an average age of 35.86 years and a median of 35 years. This diverse age profile indicates the sample’s broad reach across different population groups affected by the disease. Regarding the time between symptom onset and sample collection, the median was two days, with an interquartile range of one to three days. This finding is crucial, as most samples were obtained during the early stage of infection, a period that coincides with the peak of viremia and maximum NS1 antigen synthesis, as recommended by the World Health Organization (WHO, 2009) and described by Peeling et al. (2010).

The LABZOO laboratory protocol established RT-qPCR as the reference method for etiological confirmation and viral serotype identification. The majority of analyzed samples (92.2%) showed positive results in ELISA, while 69.2% were positive in RT-qPCR. Inconclusive results were observed in a small portion of the sample for both tests. The analysis of agreement between ELISA-NS1 and RT-qPCR revealed that 67.24% of samples showed concordant results, while 32.76% exhibited discordance between the methods. This discordance rate, where almost a third of samples reactive for NS1 were negative in RT-qPCR, demands in-depth interpretation.

The low concordance observed may be attributed to the difference in the structural stability of the biological targets. The NS1 antigen, a viral glycoprotein, has greater robustness and persists in the bloodstream for longer periods than viral RNA (Erra et al., 2013; Tricou et al., 2011). Viral RNA, in turn, is intrinsically unstable and susceptible to degradation by endogenous RNases and thermal variations during transport and storage (Bessoff et al., 2008). Thus, RT-qPCR detection is strictly linked to the active viral replication phase, which declines rapidly after the third day of symptoms, while NS1 can remain detectable even when viral RNA has already become undetectable by molecular methods (Peeling et al., 2010).

The half-life of the NS1 protein in serum is longer than that of the viral genetic material, which allows the measurement of the antigen by ELISA in samples collected between the fourth and fifth day of symptoms, when the RNA may already be undetectable (Alcon et al., 2002; Muller et al., 2017). Thus, divergent results do not necessarily indicate poor performance of ELISA-NS1, but rather reflect the assay’s ability to detect residual antigens in stages of infection in which viral RNA is no longer detectable. This distinction in the detection window of each biomarker is fundamental for understanding the discrepancies.

The package insert of the ELISA-NS1 kit from the manufacturer Bioclin reports a clinical sensitivity of 96% and specificity of 96.3%, based on a restricted panel of 147 samples and using another ELISA-NS1 assay as a reference. However, the present study, by evaluating a significantly larger sample universe (1,176 samples) and using RT-qPCR as the gold standard, demonstrated an observed agreement of only 28.7% with RT-qPCR. This discrepancy highlights the importance of validations in routine laboratory scenarios and with more robust reference methods, such as RT-qPCR, for a more accurate assessment of diagnostic performance under real-world usage conditions.

The interpretation of the divergence between the manufacturer’s results and the findings of this study should consider the intrinsic limitations of the initial kit validation and the dynamics of the host immune response. Although the NS1 antigen can be detected between the first and ninth day after symptom onset, its concentration in serum decreases with the progression of the infection. In secondary infections, the rapid formation of immune complexes between antibodies and the NS1 antigen can mask the antibody binding sites in the assay, reducing ELISA sensitivity to less than 50% from the fourth day of symptoms (Tricou et al., 2011).

The statistical analysis of agreement between the ELISA and RT-qPCR methods, using Cohen’s Kappa coefficient, resulted in a value of 0.079 (p < 0.001). According to Landis and Koch’s criteria (1977), this value indicates a very weak agreement, suggesting that the methods are not interchangeable in this sampling. Although ELISA demonstrated high sensitivity in identifying 96.3% of positive samples in RT-qPCR, the agreement was substantially reduced in negative samples, where 83.8% of negative results in RT-qPCR were classified as reactive by ELISA. This significant percentage of divergent results justifies the low Kappa index value.

The interpretation of the Kappa index should consider its sensitivity to the prevalence of the event in the sample (Sim, 2005; Wright, 2005). In the context of this study, where ELISA-NS1 showed a marked predominance of positive results, any divergence from RT-qPCR, which is the gold standard, contributed to the reduction of the agreement coefficient. The high percentage of reactive samples in ELISA (83.8%) that were not confirmed by RT-qPCR distorts the Kappa index, even with the high nominal sensitivity of the kit. The majority of samples (75%) were collected in the first three days of symptoms, a period when both targets should be at their detection peaks, reinforcing that the divergence is not primarily due to collection time, but rather to the intrinsic analytical characteristics of the methods (Hunsperger et al., 2014; Kumarasamy et al., 2007).

The distinct vulnerability of each biological target to pre-analytical conditions is a relevant factor. RT-qPCR, although highly specific, depends on the integrity of viral RNA, which is susceptible to thermal degradation in case of cold chain fluctuations or transport delays (Fry et al., 2011; Lima et al., 2010). In contrast, ELISA-NS1, with its greater structural stability, is less sensitive to these logistical variables (Bessoff et al., 2008). However, ELISA-NS1 has intrinsic limitations related to the host immune response, such as the formation of immune complexes in secondary infections (Tricou et al., 2011; Erra et al., 2013). Maintaining both tests in the LABZOO workflow, therefore, aims to mitigate the diagnostic gaps inherent to each method, acting as a safety strategy for laboratory surveillance (Peeling et al., 2010).

The analysis of the diagnostic tests’ agreement in relation to the epidemiological variables did not show a significant difference regarding the participants’ sex (p=0.584), with a similar distribution between men and women in the concordant and discordant groups. However, a significant association was observed between the tests’ agreement and the month of sample collection (p<0.001). Concordant results were mainly concentrated in March and April, representing approximately 29.5% and 33.6% of the agreement, respectively. In contrast, the agreement was null between August and October. This seasonal variation, although it may be influenced by the reduced sample size in months of low incidence, is consistent with the epidemiological dynamics of dengue.

During the first semester, the exponential increase in suspected dengue cases in the municipality of São Paulo resulted in a more representative sample of viral circulation, which tends to stabilize agreement indices (Altman and Bland, 1994). On the other hand, from August to October, the low incidence of dengue and the reduced number of samples processed by the laboratory make the statistical coefficients more susceptible to individual variations (Sim and Wright, 2005). The lack of agreement observed in the second semester reflects not only a possible reduction in the circulating viral load but also the statistical limitation imposed by the low sample density characteristic of periods with low incidence of dengue cases (Porto, 2014; Guzman et al., 2015).

The agreement between ELISA-NS1 and RT-qPCR proved to be sensitive to the epidemiological dynamics of dengue in the municipality (p < 0.001). During peak months (March and April), the convergence between methods was maximal, a phenomenon explained by the increase in Positive Predictive Value (PPV) in high prevalence scenarios (Altman and Bland, 1994). In contrast, the null agreement observed between August and October suggests that, with the reduction in viral circulation and the consequent drop in PPV, the relative impact of false-positive results (detection of NS1 antigens without active viral replication) becomes proportionally greater. This highlights the direct dependence between the analytical performance of ELISA-NS1 and the prevalence of the virus in the population (Porto, 2014; Guzman et al., 2015).

A significant difference in participant age was observed between the groups with concordant and discordant results (p<0.001). Individuals with concordant results presented a higher mean age (38.31 ± 18.07 years; median of 37 years) compared to individuals with discordant results (32.03 ± 19.4 years; median of 30.5 years). However, there was no significant difference in the number of days between symptom onset and sample collection between the two groups (p=0.311), with similar medians of approximately two days in both groups.

Although studies associate method agreement with synchrony between viremia peak and NS1 synthesis (Pillay et al., 2025), the results of this study suggest that symptom duration was not the determining factor for the convergence of results. The absence of a significant difference in sample collection interval between age groups indicates that agreement is more influenced by the susceptibility of the virus’s genetic material to pre-analytical factors and by host-intrinsic factors, such as age (p<0.001), than strictly by infection time (Fry et al., 2011; Lima et al., 2010). The maintenance of the dual protocol at LABZOO is, therefore, fundamental to ensure diagnostic sensitivity.

The finding that patients with concordant results present a significantly higher age range suggests that host-intrinsic factors modulate the detection kinetics of biological targets. In older individuals, the higher probability of previous exposure to multiple serotypes, characterizing secondary infections, may result in a distinct immunological profile (Perdomo-Celis et al., 2017; Verhagen, 2014). This immune memory response reacts immediately, combating the infection more effectively and altering the infection dynamics, reducing viremia time and the persistence of viral RNA and NS1 antigen in the bloodstream (Peeling et al., 2010).

It is plausible that, in this older population, the detection period of both markers tends to coexist within a narrower interval, narrowing the diagnostic windows and increasing the agreement between RT-qPCR and ELISA-NS1. Alternatively, the greater tendency towards immunosuppression in older populations may favor viral replication (Verhagen, 2014), resulting in a higher viral RNA load and NS1 concentration, facilitating simultaneous detection by both methods and, consequently, a higher agreement than observed in younger individuals.

The intensity of the reaction in the ELISA test, expressed by spectrophotometric absorbance, also showed a significant association with test agreement (p<0.001). Samples with concordant results exhibited higher absorbance values when compared to discordant samples. The median absorbance was 1.1 for concordant cases and 0.4 for discordant ones, suggesting that tests with significantly expressive absorbance values in ELISA tend to be positive in RT-qPCR. This result indicates that the low statistical agreement (Kappa Index) is driven by samples with borderline reactivity.

Low intensity signals near the cutoff value may reflect both the decline phase of viral load and analytical limitations associated with non-specific reactions (Bosch et al., 2017; Tricou et al., 2010). It is recommended that results with absorbance values close to the detection limit be validated by complementary methods, such as the search for IgM antibodies after the sixth day of symptoms. This approach is particularly important during epidemiological periods of low incidence, aiming to mitigate the risk of false-positive results for ELISA-NS1 and provide greater certainty in the case closure of dengue (Blacksell et al., 2008; Guzman et al., 2010).

When evaluating the relationship between ELISA absorbance values and RT-PCR classifications (Positive, Negative, and Inconclusive), statistically significant differences (p<0.001) were observed. Samples with positive RT-qPCR showed significantly expressive absorbance values in the ELISA test when compared to samples with negative RT-qPCR. Post-hoc analyses indicated that these differences occurred specifically between the negative and positive RT-qPCR groups, while comparisons involving inconclusive results in ELISA did not show statistically significant differences. These results suggest that tests with more expressive absorbance values in ELISA tend to be confirmed by RT-qPCR.

This trend suggests a correlation between viral RNA load and NS1 antigen synthesis (Alcon et al., 2002; Tricou et al., 2010). However, the presence of samples with high ELISA absorbance, in contrast to negative RT-qPCR results, reinforces the hypothesis that pre-analytical factors, such as early viral RNA degradation versus greater biochemical stability of NS1, favor the occurrence of discrepancies. Additionally, the patient’s clinical evolution time and the interference of endogenous components, such as a less effective immune response or latency in antigen recognition, also contribute to these divergences (Hunsperger et al., 2009). The correlation between biological and methodological variables confirms the need for a contextualized interpretation of results, considering the intrinsic vulnerabilities of each target, especially in a context where monitoring pre-analytical steps is challenging.

Among the samples with a positive RT-qPCR result, a predominance of the DENV-2 serotype was observed, accounting for 80.9% of the identified cases. The DENV-3 and DENV-1 serotypes were detected in considerably smaller proportions, while a small portion of the samples yielded inconclusive results for serotype determination. This predominance of DENV-2 is relevant for the interpretation of concordance results, as the sensitivity and analytical performance in NS1 antigen detection can vary significantly depending on the circulating serotype (Pillay et al., 2025). The low frequencies of DENV-1 and DENV-3 preclude a consistent analysis of the serotype’s influence on test concordance (Guzman et al., 2015).

When evaluating the relationship between the identified serotypes and the ELISA test result, it was found that most samples with a defined serotype showed a positive result in the ELISA. Inconclusive or negative results in the ELISA were observed only in samples infected by DENV-2, although this difference did not reach statistical significance in Fisher’s exact test (p=0.171). These results suggest that any discrepancies between the diagnostic methods occurred mainly in DENV-2 infections, possibly due to the higher frequency of this serotype in the analyzed sample.

The comparison of Ct values between the different serotypes resulted in a statistically significant difference (p<0.001). Samples from individuals infected by DENV-1 showed lower Ct values (median 21.7) when compared to infections by DENV-2 and DENV-3 (medians close to 26), indicating a statistically more expressive average viral load in the DENV-1 group. The higher viral load observed in the group of individuals carrying DENV-1 agrees with 100% of the positive results in the ELISA-NS1 test. In contrast, less expressive viral loads observed mostly in DENV-2 suggest that the results obtained for this serotype are close to the analytical sensitivity thresholds in both assays.

The reduced viral load presented by individuals carrying DENV-2, compared to DENV-1, may be attributed to the dynamics of secondary infections in the São Paulo population. In individuals previously exposed to other serotypes, the secondary immune response can promote a faster attenuation of viral replication, maintaining viremia at lower levels during the acute phase (Guzman et al., 2015; Thomas et al., 2010). In this scenario of reduced viral load, the simultaneous detection of biological targets in both tests tends to decrease, impacting the agreement of the results. The lower average viral load obtained in RT-qPCR for DENV-2 justifies the predominance of positive results, but with absorbance values close to the cutoff point in ELISA-NS1.

The differences in positivity between the ELISA-NS1 test and RT-qPCR were evaluated throughout the Epidemiological Weeks (EW). Both laboratory methods were capable of capturing the trends of rise, peak, and decline of the dengue epidemic in the municipality. However, the curve corresponding to the ELISA results remained systematically higher than the RT-qPCR curve throughout the monitored period. The area between the curves represents the volume of reactive samples in the ELISA test that were not confirmed by RT-qPCR. This gap reflects the higher positivity rate of the ELISA assay when compared to the RT-qPCR assay under conditions associated with laboratory routine (Datta, 2010).

The greater stability of the NS1 antigen, compared to the greater susceptibility of viral RNA degradation to variables associated with the time elapsed between collection and processing, in addition to storage conditions, contributes to the antigen detection presenting a stable diagnostic window of opportunity that is less sensitive to pre-analytical oscillations than RNA detection methods, such as RT-qPCR (Ogire et al., 2023). This biological characteristic of NS1 is a determining factor for the LABZOO routine, ensuring consistent detection performance within the recommended time window of zero to five days.

Spearman’s correlation analysis between laboratory variables showed a significant positive association between RT-PCR Ct values and the number of days since symptom onset (r = 0.367; p < 0.001). This result reflects a progressive increase in Ct throughout clinical evolution, which translates to a gradual reduction in viral load as the infection progresses. This dynamic is consistent with the dengue viremia kinetics described by Peeling et al. (2010) and Guzman et al. (2010), in which the peak of viral replication is concentrated in the initial days, followed by a sharp decline mediated by the host’s immune response.

The correlation between ELISA absorbance and symptom onset time also showed a discrete, but statistically significant variation (r = 0.062; p = 0.039). Such correlation suggests that the absorbance values among the positive results in the ELISA test exhibit discrete variation throughout the clinical evolution, evidencing a stability of the NS1 antigen superior to that observed in viral RNA. These results converge with the studies by Erra et al. (2013) and Teoh et al. (2016), which demonstrated that the half-life of the NS1 protein in serum is considerably longer than that of the viral genetic material.

A negative correlation was also observed between RT-qPCR Ct values and absorbances obtained in the ELISA test (r = -0.132; p < 0.001). Although statistically significant, this correlation showed a discrete magnitude in the analyzed sample universe. The negative coefficient confirms that samples with higher viral load (lower Ct values) tend to present higher absorbance values in ELISA. The insignificant value associated with this correlation suggests, however, that the absorbance value does not have a strictly linear relationship with the viral RNA load present in the individual. The stability of the NS1 antigen in the serum sample is independent of the integrity of the viral genetic material, which justifies the partial dissociation observed between the methods (Duyen et al., 2011; Tricou et al., 2011).

IgM class antibodies represent the primary immune response to dengue virus infection, with detection typically viable from the sixth day after symptom onset. At this stage, the IgM antibody acts as a strategic target for confirming cases in which viral load and NS1 antigen are already declining due to the humoral response (Teoh et al., 2015). In the context of this study, the ELISA-IgM assay plays a complementary role in closing dengue cases with negative or borderline results in ELISA-NS1 and RT-qPCR tests, increasing diagnostic sensitivity by covering the different biological windows of the disease (Erra et al., 2013; Blacksell et al., 2011; Fry et al., 2011; Teoh et al., 2016).

Considering that LABZOO already has an established workflow for performing the ELISA-IgM test on samples collected from the sixth day of symptoms (Executive Secretariat of Primary Care [SEABEVS], 2025), the adoption of an integrated laboratory strategy is recommended. For patients with clinical suspicion of dengue who present undetectable results in ELISA-NS1 and RT-qPCR in early samples, active surveillance or scheduling a new collection after the sixth day of symptom onset for IgM antibody research is advised. Although anticipating collection to the fourth or fifth day may be considered, the sixth day remains the benchmark of greatest diagnostic certainty for antibody detection, minimizing the risk of false-negative results due to low titers (Peeling et al., 2010). The systematic articulation between ELISA-NS1, RT-qPCR, and ELISA-IgM assays, contextualized to the epidemiological scenario of São Paulo, will allow for the mitigation of the intrinsic limitations of each method, improving dengue laboratory surveillance in the municipality.

In summary, the results of this study showed low agreement between ELISA-NS1 and RT-qPCR assays in the routine of LABZOO, influenced by epidemiological, laboratory, and clinical variables. The greater stability of the NS1 antigen compared to viral RNA and the dynamics of the host’s immune response, especially in older age groups, contributed to the observed discrepancies. The sensitivity of ELISA-NS1 proved to be dependent on viral prevalence and absorbance intensity, while the predominance of the DENV-2 serotype and its lower average viral load were associated with results close to the detection limits. These findings reinforce the need for an integrated diagnostic workflow and rigorous analytical criteria in kit acquisition to optimize laboratory surveillance of dengue in the municipality of São Paulo.

4. Conclusion

This study evaluated the statistical agreement between ELISA-NS1 (Bioclin Kit) and RT-qPCR assays, used in the routine of the Zoonoses Laboratory (LABZOO) of the municipality of São Paulo, from January to October 2025. Low agreement was found between the methods, influenced by epidemiological, laboratory, and clinical variables. It was observed that the agreement varied according to dengue seasonality, being higher in months of epidemic peak (March and April) and null in periods of low incidence (August to October). Individuals of older age groups showed greater agreement, and concordant results exhibited higher absorbance values in ELISA and lower Ct values in RT-qPCR, indicating a higher viral load. The NS1 antigen demonstrated greater stability over the time of symptoms compared to viral RNA, which contributed to the higher positivity rate of ELISA compared to RT-qPCR. The predominance of serotype DENV-2, with less expressive viral loads, suggested that the results were close to the analytical sensitivity thresholds. These findings provide crucial technical subsidies for optimizing the laboratory flowchart for dengue diagnosis in the municipality of São Paulo.

The study presented limitations inherent to its retrospective nature and the comparison of methodologies that detect distinct biological targets, in addition to the impossibility of monitoring pre-analytical conditions and the absence of intra-laboratory analytical validation. Therefore, for future studies, a comparative evaluation between commercial NS1 antigen kits from different manufacturers is recommended, seeking inputs with greater sensitivity and concordance with RT-qPCR. The implementation of rigorous internal quality control protocols is also suggested, including the validation of repeatability and reproducibility, and the pre-validation of kit batches in reference laboratories. It is proposed that future bidding processes for the acquisition of ELISA kits incorporate rigorous analytical performance metrics. Finally, the need to consolidate an integrated diagnostic workflow between RT-qPCR, ELISA-NS1, and ELISA-IgM at LABZOO is reinforced, allowing for an expanded detection window and mitigating the intrinsic limitations of each method, ensuring more consistent diagnoses for dengue surveillance.

Bibliographic References

Altman, D.G.; Bland, J.M. 1994. Diagnostic Tests 2: Predictive Values. BMJ (Clinical Research Ed.) 309(6947): 102.

Castro-Jorge, L.A.; Machado, P.R.L.; Fávero, C.A.; Borges, M.C.; Passos, L.M.R.; De Oliveira, R.M.; Fonseca, B.A.L. 2010. Clinical Evaluation of the NS1 Antigen-Capture ELISA for Early Diagnosis of Dengue Virus Infection in Brazil. Journal of Medical Virology 82(8): 1400–1405.

COVISA – Coordenação De Vigilância Em Saúde – Município De São Paulo. 2025. Boletim arboviroses. Diário Oficial da Cidade, São Paulo.

CVE São Paulo – Centro de Vigilância Epidemiológica. 2025. Situação Epidemiológica Arboviroses Urbanas: Dengue, Chikungunya e Zika. Relatório Técnico, São Paulo, SP, Brasil.

Duyen, H.T.L.; Ngoc, T.V.; Ha, D.T.; Hang, V.T.T.; Kieu, N.T.T.; Young, P.R.; Farrar, J.J.; Simmons, C.P.; Wolbers, M.; Wills, B.A. 2011. Kinetics of plasma viremia and soluble nonstructural protein 1 concentrations in dengue: Differential effects according to serotype and immune status. Journal of Infectious Diseases 203(9): 1292–1300.

Erra, E.O.; Korhonen, E.M.; Voutilainen, L.; Huhtamo, E.; Vapalahti, O.; Kantele, A. 2013. Dengue in Travelers: Kinetics of Viremia and NS1 Antigenemia and Their Associations with Clinical Parameters. PLoS ONE 8(6).

Guzman, M.G.; Jaenisch, T.; Gaczkowski, R.; Hang, V.T.T.; Sekaran, S.D. [et al.]. 2010. Multi-country evaluation of the sensitivity and specificity of two commercially-available NS1 ELISA assays for dengue diagnosis. PLoS Neglected Tropical Diseases 4(8): 2-11.

Instituto Brasileiro de Geografia e Estatística, 2025 [Referência completa não encontrada no documento original]

Lorenz, C.; Lemey, P.; Dellicour, S. 2026. Dengue serotypes and epidemic dynamics in Brazil: a spatiotemporal perspective. Travel Medicine and Infectious Disease 69(September): 102948.

Pillay, K.; Keddie, S.H.; Fitchett, E.; Akinde, C.; Bärenbold, O. [et al.]. 2025. Evaluating the performance of common reference laboratory tests for acute dengue diagnosis: a systematic review and meta-analysis of RT-PCR, NS1 ELISA, and IgM ELISA. The Lancet Microbe 6(7): 1–8.

SEABEVS – Secretaria Executiva de Atenção Básica, Especialidades e Vigilância em Saúde. 2025. Orientações para o Diagnóstico Laboratorial das Arboviroses. Relatório Técnico, São Paulo, SP, Brasil.

Teixeira, M.A.G.L. 2008. Epidemiological situation and dengue control in Brazil. Revista USP 22(64): 53–72.

Teoh, B.T.; Sam, S.S.; Tan, K.K.; Johari, J.; Abd-Jamil, J.; Hooi, P.S.; Abubakar, S. 2016. The Use of NS1 Rapid Diagnostic Test and qRT-PCR to Complement IgM ELISA for Improved Dengue Diagnosis from Single Specimen. Scientific Reports 6(March): 1-8.

Vicente et al., 2016 [Full reference not found in the original document]

Article originating from the Final Course Project of the Specialization in Data Science and Analytics of the MBA USP/Esalq

To learn more about the course, click here and access the MBX Academy platform

You may also like

Executive Mba In Leadership And Management

November 23, 2026

Improvements in the efficiency of a secondary biological sludge drying plant

The management of industrial solid waste, especially secondary biological sludge, represented an operational and strategic challenge for a company in the pulp and paper sector. A secondary biological sludge drying plant, installed to meet environmental requirements and reduce costs, operated with low efficiency and high unavailability. The objective of this study was to evaluate and propose a set of actions to increase the operational availability of the plant, ensure the integral processing of the generated sludge, and reduce treatment and transportation costs, in addition to mitigating environmental risks. To this end, an analysis of the root causes of low efficiency was carried out, using tools such as the cause-and-effect diagram (Ishikawa) and the “5 whys” analysis, and hypotheses were tested to identify the most impactful factors. Three main root causes were identified: addition of low-load effluent to the process, absence of a dedicated maintenance team, and lower consistency of dewatered sludge. The proposed countermeasures, which included stopping the addition of effluent, hiring a maintenance team, and testing polymers, had a monthly implementation cost of R$ 35,000.00. An annual reduction in operating expenses of up to R$ 1,800,000.00 and maintenance costs of R$ 600,000.00 was estimated, in addition to a 25% reduction in operational unavailability. It was concluded that the proposed measures are viable, quick to implement, and capable of ensuring the integral processing of secondary biological sludge, strengthening the company’s ESG strategy and ensuring environmental compliance.

Keywords: Operational costs; Operational efficiency; ESG; Biological sludge; Sludge drying.

Data Science And Analytics For Operations

November 19, 2026

Polyp-YOLO: an efficient, low-latency detector for polyps in colonoscopies

The early detection of colorectal polyps is crucial for the prevention of colorectal cancer, and real-time computer-aided detection (CADe) systems have shown gains in endoscopic practice. The objective was to develop and validate a lightweight colorectal polyp detector, based on the YOLO architecture, capable of operating on restricted hardware with high precision and low latency. The methodology used public datasets of endoscopic images, partitioning the Kvasir-SEG dataset for training and internal validation, and CVC-ClinicDB and ETIS-Larib Polyp DB for cross-dataset evaluation. Lightweight variants of the YOLOv8, YOLO11, and YOLO26 families with 640×640 pixel input were evaluated. A grid search tested 70 combinations of optimizers and activation functions, and identified AdaBelief and LeakyReLU as the configuration with the best generalization. The optimization pipeline applied unstructured pruning, knowledge distillation, and fine-tuning. The optimized YOLO11n-seg model outperformed the higher-capacity architecture (YOLO26m-seg) in mAP50 and showed similar performance in F1-Score, operating with approximately seven times fewer parameters. It achieved an mAP50 of 0.7929 and an F1-Score of 0.7514 on the ETIS-Larib dataset, and fast inference was demonstrated in different formats, with real-time latencies. The explainability of predictions was validated by Grad-CAM activation maps. The proposed solution demonstrated itself to be a promising experimental tool for diagnostic support, with technical feasibility for integration into environments with reduced computational power, although it requires future clinical validation.

Keywords: Computer-aided detection; Edge computing; Hyperparameter optimization; Endoscopic image segmentation; Medical computer vision.

Data Science And Analytics For Operations

November 19, 2026

Multi-touch Attribution in E-commerce: A Comparison between Markov Chains and Last-Click

The research investigated Multi-touch Attribution (MTA) Models in e-commerce, seeking to overcome the limitations of the Last-Click model, widely used in the market. The objective was to compare the performance of first-order Markov Chains with Last-Click and Position-Based approaches, identifying the value of assistance channels frequently ignored. To this end, Markov Chain algorithms were applied to a large volume of purchase journey data from the Google Merchandise Store, extracted via BigQuery. The study probabilistically mapped the purchase journey and calculated the Removal Effect to quantify channel relevance. The results demonstrated that data-driven models identified the strategic role of discovery (awareness) and assistance channels, which Last-Click neglected in 65% of complex journeys. Simulated financial analysis indicated an approximate 28% hidden efficiency in Cost Per Acquisition (CPA) for Paid Search, compared to the Last-Click model. It was concluded that Markov Chains allowed for a more accurate mapping of purchase journeys, validating the technical feasibility of the modeling and providing evidence for more efficient budget allocation in digital media, mitigating the risk of early divestment in crucial channels.

Keywords: Assistance channels; Data science; E-commerce; Digital marketing; Media optimization.

Project Management

November 19, 2026

Integrated Software for Operational Communication and Control: Practical Evidence in a Medium-Sized Company

The study analyzed the impacts of the lack of integration between commercial and operational modules of an integrated management software (ERP/CRM) in a medium-sized company in the mechanical engineering sector. The main objective was to evaluate practical evidence of communication and operational control optimization, identify bottlenecks, and formulate improvement proposals. A qualitative-quantitative approach was adopted through a single case study with documentary research, based on the extraction and filtering of historical task records from the Bitrix24 software and statistical correlation analysis between workload and allocated team. Based on the diagnosis, an action plan structured in the 5W2H tool was developed. The results revealed that 54.4% of the registered tasks consisted of duplicate, test, or practically useless records, generated by manual insertion during the scope transition between commercial CRM and engineering. It was also demonstrated that the increase in the commercial team linearly accompanied the growth of inefficient tasks, polluting the database and masking the organization’s real productive capacity. It was concluded that the automation of workflows and native integration between modules are essential to mitigate operational and financial risks, providing replicable guidelines for data governance in small and medium-sized enterprises.

Keywords: System Customization; Enterprise Resource Planning (ERP); Task Management; Management Software.

Project Management

November 19, 2026

Supervised learning in voltammetric quantification with electrochemical sensors

The application of supervised learning in the voltammetric quantification of paraquat in artificial urine was evaluated. The study aimed to quantitatively assess the predictive performance of supervised learning models compared to a univariate “baseline”, preserving independence between devices in the training and testing sets. The research was conducted as an applied case study, processing data from four physical devices, each evaluated under a blank condition and six paraquat concentration levels, totaling 28 square wave voltammetry (SWV) curves. A computational “pipeline” integrated automated file reading, organization of curves into current-potential matrices, generation of potential maps, and device-specific blank correction. Five approaches were considered: conventional univariate calibration, predictive univariate “baseline”, partial least squares (PLS) regression, support vector regression (SVR), and random forest (RF). Predictive evaluation was performed using device-grouped cross-validation. PLS showed the best predictive performance for curves from a device not used in training, with an R² of 0.7981. RF exhibited intermediate performance, while SVR and the predictive univariate “baseline” showed inferior and similar performance between themselves. It was concluded that a compact multivariate representation of the voltammetric profile was more favorable for inter-device prediction than increasing model flexibility under the studied conditions. Grouped validation preserved independence, but the limited number of devices restricts the generalization of results.

Keywords: Supervised learning; Paraquat; PLS; Electrochemical sensors; Square wave voltammetry.

Project Management

November 19, 2026

Improvement of Inventory Management via Lean Six Sigma: Study in the Paper Industry

Organizations seek methodologies to reduce waste, increase efficiency, and become more competitive, promoting sustainable improvements in their processes. In this context, the study aimed to apply the Lean Six Sigma methodology, through the DMAIC cycle, to improve inventory management in a company in the paper industry manufacturing pneumatic hoses, aiming to increase the accuracy between physical and systemic inventories and improve the visual management of parts storage. The research was developed from the collection and analysis of data extracted from the company’s ERP system. Tools such as SIPOC, Ishikawa Diagram, and 5 Whys were used to define and identify the root cause of the problem. As improvement proposals, a Digital Kanban system and an adaptation of the Two-Bin system, called the Three-Bin System, were implemented, in addition to the development of standardized operating procedures and the establishment of control mechanisms. The results demonstrated a significant improvement in inventory accuracy, raising the lowest identified index from 28% to 80%, and detected discrepancies between systemic and physical records equivalent to R$ 134,335.53. It was concluded that the application of Lean Six Sigma, combined with Project Management, represented an effective approach for conducting operational improvement projects, contributed to increased reliability of inventory information, to the improvement of material control processes, and to the strengthening of the organization’s operational management.

Keywords: Inventory accuracy; DMAIC; Visual management.

Data Science And Analytics For Operations

November 19, 2026

Predictive Analysis Methodology for Performance Optimization in E-commerce: Correlation between Public Data and Paid Media

A predictive analysis methodology was developed for e-commerce, integrating transactional data and public search trends. The objective was to develop and validate a predictive model, based on machine learning algorithms, capable of correlating organic search trends with sales performance and paid media costs in the Beauty and Health category of Brazilian e-commerce. The Brazilian E-Commerce Public Dataset by Olist and data from Google Trends were used, processed in Python with the Random Forest Regressor algorithm. A moderate inverse correlation (r = -0.43) with a seven-day lag was identified between organic search volume and purchase. Initial models, based solely on behavioral variables, presented insufficient metrics (R² = -0.47; MAE of R$ 2,120.54). The incorporation of microeconomic (average ticket and shipping) and macroeconomic (exchange rate and Ibovespa index) variables raised the R² of the final model to 0.29, with an MAE of R$ 1,489.63, representing a 24.7% reduction in error compared to the baseline. Temporal cross-validation showed that, in periods of mature operation (2018), the model achieved an R² of 0.42. It was concluded that isolated search volume explains the moment of consumer interest, but not the financial magnitude of sales, making the fusion with transactional and economic data mandatory for predictive viability.

Keywords: Beauty and Health; Consumer Behavior; Google Trends; Machine Learning; Time Series.